Sign In | Join Free | My ccnmag.com |
|
Brand Name : Sinuo
Model Number : SN9620X
Certification : Calibration Certificate (Cost Additional)
Place of Origin : China
MOQ : 1
Price : Negotiable
Payment Terms : L/C, T/T, Western Union, MoneyGram
Supply Ability : 50 sets per month
Delivery Time : 20 Days
Packaging Details : Carton Box
Test Contents : Leakage current test
Applied Standard : IEC60335.1, UL60335.1, etc.
Test Input Voltage : 30-300V AC Single phase
Test Input Current : 20A AC MAX
Leakage Current Range : 50-6000uA
Leakage Current Resolution : 1uA
Leakage Current Tester Electrical Appliance Safety Testing Equipment
Product information:
■ Meet the electrical safety general test requirements of IEC60335.1, UL60335.1, GB4706.1, etc.
■Built-in common human body simulation impedances, different simulation impedance standard cards can be selected according to different standards, default come with network as per IEC 60335-1, other networks are optional.
■ Upper limit preset, automatic reminder of alarm message, intelligent memory analysis function, system self-check function, automatic protection function.
■ 16*2LCD display
■The instrument has a software calibration function for simplify parameter calibration.
Technical Parameters:
Model | SN9620X | |
Input voltage | 220Vac±10% | |
Input power | 50/60Hz±5% | |
Fuse | Fuse 0.5A/250V | |
Leakage current | ||
Test input voltage | 30-300V AC Single phase | |
Test input current | 20A AC MAX | |
Voltage display range | 30-300V AC | |
Voltage display resolution | 1V | |
Voltage display accuracy | ± (2% setting+3V) | |
Leakage current range | 50-6000uA | |
Leakage current resolution | 1uA | |
Leakage current accuracy | ± (2% reading+30uA) | |
Built-in human network | Default as per IEC 60335-1, other networks are optional | |
Test time range | 1.0-2.0-999.9s , resolution 0.1s, set 1s only to test the leakage of firewire | |
Function parameters | ||
Display | 16*2 blue screen character LCD display | |
Communication Interface | RS232(optional) | |
Calibration method | Software calibration | |
Working environment | 0-40℃, 20-80%RH | |
Dimensions & weight | 280mm(W)×89mm(H)×370mm(D), approx. 15kgs | |
Accessories | Power cord × 1, test clip × 2, fuse× 1 |
Calibration required instrument:
1. 50-300V adjustable test power source
2. 50K block resistor
3. Digital multi-meter with a 20mA AC.
Calibration preparation:
Set the upper limit current alarm value of the tester SN9620X to the full scale, and the lower limit to 0. The current calibration within the measuring range can be performed and avoid the over current test stop.
Calibration connection:
![]() |
Leakage Current Tester Electrical Appliance Safety Testing Equipment Images |